Catheters in the Root - Engaging the Coronaries, Step by Step
Push, pull and turn coronary catheters in one patient's aortic root, from the groin or either wrist, and see what each move does to the tip, with a coach that gives the next move for the catheter in your hand.
Every coronary catheter is engaged with four moves at the hub: push, pull, clockwise and anticlockwise. What each one does to the tip depends on the curve, on the route the catheter took to get there and on the root it sits in, and in the lab that is learned slowly, one patient at a time. Here the catheter sits in one patient’s aorta and aortic root, with the three cusps and the coronary ostia, and every move at the hub shows on the tip at once.
The catheters bend and twist against the walls. A turn winds into the shaft before it reaches the tip, a push takes up the slack first, and a tip held in a cusp stores the turn until it lets go. A Judkins left drops into the left main by itself from the groin, and lands in the right cusp from the wrist. The pressure at the tip shows the aorta, the ventricle, or a damped trace when the end hole is against the wall.
Using it in a teaching session
Start from the groin with a Judkins left and a Judkins right: up over the arch on the wire, the wire out, and into each ostium. Then do the same from the right wrist, where the same curve sits differently. Take a prediction before each move: which way the tip will go, and how far.
Switch on the coach and a line under the picture gives the next move for the catheter in hand, from where its tip is now, and lights the control that makes it: two at once when a turn and a push or pull go together. It names the engaged ostium and says when to inject, when the tip is too deep or the pressure is damped, and when a curve is the wrong size for the root.
Nineteen demonstrations show the moves an operator would make, from the way up from the groin and from either wrist, through a loop in the subclavian, to the Amplatz curves, extra-backup and radial guides, the pigtail across the valve and engaging the left main through a TAVI frame. Practise gives a random access, catheter and target, and scores the time, the contrast and the safety.
The patient can be changed: a small, normal or dilated root, separate LAD and circumflex ostia, a circumflex from the right sinus, an RCA from the left sinus or high and anterior, right, left or co-dominance and a ramus, severe aortic stenosis, and self-expanding or balloon-expandable TAVI valves. The views include AP, LAO, RAO, lateral and down the root from above, and each projection can be seen as fluoroscopy, dark until contrast fills the root.
What the model simplifies
The aorta and coronary arteries come from one patient’s CT. The catheter curves are generic shapes: each maker’s version differs a little, and the friction and stiffness are typical rather than those of any one catheter. The heartbeat, the contrast and the pressures are illustrative. The coach reads the tip the way an operator reads the screen; it is guidance, not a rule. It teaches catheter handling and is not for clinical use.
Sources
Catheter selection and technique follow Kern’s Cardiac Catheterization Handbook (Sorajja, Lim and Kern, 7th ed., Elsevier 2020) and the AHA statement on transradial access (Mason et al., Circ Cardiovasc Interv 2018;11:e000035). Coronary access after TAVI follows Yudi et al. (J Am Coll Cardiol 2018;71:1360-78), and the coronary anomalies, Angelini (Circulation 2007;115:1296-305).