The C-arm and Fluoro Trainer - Table, Detector, Collimation and Dose
A table-side trainer for cath lab technicians: drive the C-arm, the table and the detector at real speeds, with collision sensors, and watch every move on a live fluoroscopy image and on its dose.
Most of the radiation a patient and a team receive in the cath lab is decided at the table side, by where the table, the detector and the shutters are put, before anyone reads a picture. This trainer puts those controls in a technician’s hands. On the left the C-arm, the table and the draped patient move as they are driven; on the right is the fluoroscopy image they produce, computed through a real chest CT, with the dose it costs.
The patient starts as placed, a few centimetres low and off the isocentre, as a patient usually lies before the table is set. Nothing re-centres the heart but the trainee.
Using it in a teaching session
Start with the table. Centre the heart in AP, then swing to a steep LAO and watch the heart drift across the field until the table height is right. The line at the foot of the C-arm view names the one thing to fix first: the heart off centre, the heart too close to the tube, or the detector too far from the patient.
Bring the detector down until its sensor slows it, and read the dose before and after. Then raise the detector and lower the table to see the opposite. Take a prediction before each move: which way the picture will go, and whether the dose will rise or fall.
Drive to the standard views with the joystick, then with stored positions and Accept, which moves the C-arm only while the key is held and stops when a sensor trips, as in the lab. At a steep angle with the table high, the trainee has to clear the patient before the C-arm will go on.
Set the shutters and wedges on the last image with no exposure, then compare the dose-area product and the scatter at the operator’s position with the field open and closed. The Drills page sets positioning tasks that tick themselves as they are done, and Team brings back the whole room, with the operator, the scrub assistant and the nurse at work.
What the model simplifies
Travel and speeds are typical of a floor-mounted cardiac C-arm rather than those of any one system: isocentre 106 cm above the floor, tabletop 77.5-110 cm, SID 90-120 cm, LAO and RAO to 120° and cranial and caudal to 55°. The controls are a generic teaching layout, and a lab’s own keys, names and limits will differ.
The collision sensors know the patient’s chest, arms and head and the tabletop, against the face of the detector and the tube housing. They do not model the arc against the table, or anything else in the room, and the Show me buttons in the drills jump straight to their answer rather than driving there. Doses are illustrative, order-of-magnitude figures: the patient’s from the tube output, the SID and the field size, and the team’s scatter from the dose-area product. A laboratory’s own numbers come from its physicist.
Sources
The patient is an anonymised contrast-enhanced chest CT of a normal heart, segmented with TotalSegmentator (Wasserthal J et al., Radiology: Artificial Intelligence 2023) and the MONAI model zoo. The coronary tree is grown on this heart’s own surface and grooves. The image is formed by casting the beam through that CT, and both the image and the room are rendered with three.js.