CardioRender · Dr PDR's Teaching Series

Electrode to ECG

lead reversal, artifact and filters
100%
How to use this
The tracing is a clean normal sinus ECG. Cross a cable under Reversal, put noise on one electrode under Artifact, then try to remove it under Filters. Every lead an artifact reaches is flagged in orange on the paper.
Or test yourself: six tracings, each with one hidden fault to name.
Electrode to leads
RAI · II · aVR · aVL · aVF · V1–V6 LAI · III · aVR · aVL · aVF · V1–V6 LLII · III · aVR · aVL · aVF · V1–V6 RLno lead (neutral) V1–V6its own lead only
The chest leads are measured against the Wilson central terminal, (RA+LA+LL)/3, so a faulty limb electrode reaches all six of them at one third strength.
Lead distribution
Readouts
Education and training only · not a diagnostic or clinical decision tool · © Dr Dinesh Raja Palanisamy · CardioRender